| HS Code | 422280 |
| Product Name | Toluene, p.a. (pro analysi), Acros Organics |
| Grade | p.a. (pro analysi) |
| Cas Number | 108-88-3 |
| Molecular Formula | C7H8 |
| Molar Mass | 92.14 g/mol |
| Assay Gc | ≥99.5% |
| Appearance | Clear colorless liquid |
| Density | 0.865 g/cm³ at 20 °C |
| Melting Point | -95 °C |
| Boiling Point | 110-111 °C at 760 mmHg |
| Flash Point | 4 °C (closed cup) |
| Vapor Pressure | 28.4 hPa at 20 °C |
| Refractive Index | 1.496 at 20 °C |
| Solubility In Water | 0.47 g/L at 20 °C |
| Autoignition Temperature | 480 °C |
As an accredited Toluene, p.a. (pro analysi), Acros Organics factory, we enforce strict quality protocols—every batch undergoes rigorous testing to ensure consistent efficacy and safety standards.
| Packing | Toluene, p.a. (pro analysi) from Acros Organics comes in a 1 L amber glass bottle with a secure cap, ensuring purity and safe handling. |
| Container Loading (20′ FCL) | 20′ FCL loading of Toluene (p.a.) in sealed drums; flammable, toxic liquid; secure upright, grounded, away from oxidizers. |
| Shipping | Toluene, p.a. (Acros Organics) ships as a hazardous flammable liquid, UN 1294, Class 3, PG II. Packaged in approved glass or metal containers with leak-proof seals, it must be transported per IATA/IMDG/ADR regulations, away from oxidizers, with proper hazard labeling and documentation. |
| Storage | Store in a tightly sealed original container in a cool, dry, well-ventilated area away from heat, sparks, open flames, and strong oxidizers. Keep upright and protect from sunlight. Ensure proper labeling and secondary containment. Maintain room temperature, avoid unnecessary exposure, and follow local regulations for flammable liquid storage. |
| Shelf Life | Shelf life is typically 5 years when stored tightly sealed in the original container, protected from light and moisture. |
Toluene p.a. (pro analysi), Acros Organics, is specified by lot-specific certificate values for assay, water, and non-volatile residue. Release criteria commonly state assay ≥99.5% by gas chromatography, water ≤0.03% by Karl Fischer titration, and evaporation residue ≤5 ppm. Density is 0.867 g/mL at 20°C. Boiling point is 110.6°C at 101.3 kPa. The aromatic structure C7H8 imposes a UV cutoff near 285–290 nm. These values control downstream application because trace polar impurities, residual sulfur, and non-volatile matter generate interfering baselines in gas chromatography, shift refractive index response in polymer analysis, or deactivate oxidation catalysts. The following application tracks are separated by governing standard, instrument configuration, and failure threshold rather than by generic solvent use.
In pharmaceutical quality control, toluene p.a. is used primarily as a reference standard for Class 2 residual solvent determination under ICH Q3C, where the permitted daily exposure is 8.9 mg/day and the concentration limit is 890 ppm. Method execution follows USP <467> Procedure A static headspace gas chromatography with flame ionization detection. A concentrated Class 2 standard stock is prepared by dissolving 0.890 g of toluene p.a. in 100.0 mL of a high-boiling diluent such as N,N-dimethylacetamide. Working calibrators are prepared by serial dilution to cover the residual solvent range relevant to oral solid dosage forms, commonly 1/100, 1/200, and 1/500 of the stock. Headspace vials are equilibrated at 80°C for 30 min. A 1 mL headspace injection is split 5:1 onto a 30 m × 0.32 mm × 1.8 µm 624-type column. The system suitability requirement is resolution ≥1.5 between toluene and adjacent Class 2 solvent peaks, especially methylcyclohexane and cis-1,2-dichloroethene. A significant operational conflict arises from matrix retention of toluene in formulations containing polyols, polyethylene glycol, or sorbitol. Such matrices reduce headspace recovery and shift response relative to the neat standard. Matrix-matched calibration is therefore performed. The injection liner and headspace transfer line are maintained at 125°C and 130°C respectively to prevent aromatic condensation. Liner replacement after 40–50 injections is required because non-volatile formulation excipients deposit on deactivated surfaces and create active sites that adsorb toluene. Standards must be stored in glass vials with PTFE-lined septa because low-density polyethylene containers release plasticizer interferences. The terminal object is a lot-specific residual solvent report for drug product release. Samples above 890 ppm trigger out-of-specification investigation under ICH Q3C. Toluene p.a. itself is not injected into formulations; it is restricted to reference standard preparation and method verification.
When dissolved-phase concentrations fall below 1 μg/L, the limiting variable shifts from detector sensitivity to water management and carryover in the purge trap. Toluene p.a. is used to prepare methanol-based volatile organic compound calibration spikes for EPA 8260D purge-and-trap gas chromatography–mass spectrometry. The aqueous sample volume is 5 mL. Purge temperature is held at 40°C for 11 min with inert gas flow at 40 mL/min. The trap desorption is set at 180°C for 2 min. The GC column is a 30 m × 0.25 mm × 1.4 µm DB-624 with helium carrier at 1.2 mL/min. An internal standard, fluorobenzene, is added at 20 μg/L. Toluene p.a. is spiked into reagent water at 0.5 μg/L to 200 μg/L. For soil leachates, matrix spikes are prepared at 50 μg/kg wet weight. The principal failure threshold is carryover. After a 200 μg/L calibration standard, a reagent water blank must produce toluene response below 0.5 μg/L. If the blank exceeds this limit, the trap is baked at 210°C for 8 min and the water management system is inspected for condensation in the mount. Because toluene p.a. has evaporation residue ≤5 ppm, it does not contribute significant non-volatile foulants to the trap. Polar impurities in lower-grade toluene produce early column bleed and baseline rise in full-scan mode. Selective ion monitoring improves method detection limit, but published multi-laboratory data for this specific configuration vary between 0.02 μg/L and 0.1 μg/L depending on municipal water matrix and mass spectrometer tuning. The terminal use is a regulatory compliance data package for groundwater monitoring under drinking water or wastewater discharge permits. Toluene p.a. is not used as a purge vessel cleaning solvent; only blank water and methanol appropriate for purge-and-trap are applied. Open-vial storage of toluene p.a. standards in refrigerated autosamplers must not exceed 7 days because volatile loss shifts calibration response more than 5% relative to freshly prepared stocks.
| Segment | Governing standard | Instrument configuration | Toluene p.a. parameter controlling reliability |
|---|---|---|---|
| Pharmaceutical residual solvent | USP <467>, ICH Q3C | Headspace sampler at 80°C; 30 m × 0.32 mm × 1.8 µm 624-type column; FID | Evaporation residue ≤5 ppm; organic interference profile |
| Environmental VOC analysis | EPA 8260D | Purge-and-trap autosampler at 40°C; 30 m × 0.25 mm × 1.4 µm DB-624; MSD | Assay ≥99.5%; water ≤0.03% for calibration stock stability |
| High-temperature SEC | ISO 16014-4 | RI detector at 35°C; mixed-bed columns 300 mm × 7.8 mm | Water after sieve drying ≤0.01%; particulate-free filtration |
| UV-Vis additive assay | ASTM E169-16 | Double-beam spectrophotometer; 10 mm quartz cuvettes | Absorbance baseline above 300 nm; aromatic cutoff 285–290 nm |
High-temperature polymer characterization requires a mobile phase that remains chemically inert against polyolefin and polystyrene analytes in a 145°C column oven. Toluene p.a. is dried over 3 Å molecular sieves to water ≤0.01% before a 2.0 mg/mL solution of linear polyethylene is prepared. The dissolution step is conducted at 135°C for 90 min under sealed stainless-steel vials. The polymer solution is filtered through a 0.45 µm PTFE syringe filter and injected at 1.0 mL/min through a mixed-bed styrene-divinylbenzene column set with 300 mm × 7.8 mm columns. A refractive index detector maintained at 35°C records the concentration signal. For Mark-Houwink parameter generation, a viscometer and multi-angle light scattering detector are connected downstream of the concentration detector. Baseline drift above <1 × 10⁻⁶ RIU/h is traced to water ingress, dissolved oxygen, or particulate fouling of the inlet frit. The aromatic refractive index increment of polystyrene in toluene is typically 0.110 mL/g at 25°C, which improves signal-to-noise ratio compared with tetrahydrofuran. The column bank calibration is established with 12 narrow polystyrene standards from 1,000 Da to 10,000,000 Da. Retention drift between the initial and final calibration standard must remain below 0.2%. Toluene p.a. is not compatible with polycarbonate inline tubing or polycarbonate filter housings. Flow path components are restricted to stainless steel, PTFE, or fluoropolymer-elastomer seals. The terminal output is a weight-average molecular weight and dispersity report for polyethylene resin release or failure investigation. Batch-to-batch variation in sieved toluene p.a. is controlled by Karl Fischer verification before each mobile phase lot is accepted. A mobile phase lot with water above 0.015% is rejected or re-dried because refractive index baseline curvature obstructs integration of low-molecular-weight tails.
Laboratory-scale partial oxidation of toluene p.a. to benzaldehyde is conducted in a 1 L Hastelloy C276 stirred autoclave equipped with a gas entrainment impeller and baffle. The liquid charge contains 500 mL toluene p.a., 0.2 mol% cobalt(II) acetate tetrahydrate, 0.04 mol% manganese(II) acetate tetrahydrate, and 0.1 mol% sodium bromide as promoter. Air is metered at 0.4 MPa and the reactor is held at 150°C for 90 min. Conversion per pass is maintained at 15–20% to suppress consecutive oxidation to benzoic acid. Selectivity to benzaldehyde is typically 40–50% under these conditions. The remainder is benzyl alcohol, benzyl bromide, and benzyl ester intermediates. Published data for this specific autoclave charge configuration are limited; the performance range is derived from laboratory oxidation protocols rather than commercial unit operations. P.a. purity matters because trace sulfur compounds deactivate the cobalt active site. Non-volatile residues coat the impeller shaft and reduce oxygen mass transfer. If stirring speed drops below 800 rpm, conversion falls and heat accumulation triggers autocatalytic oxidation to benzoic acid. The off-gas is passed through a cold trap at -5°C to recover entrained toluene. The crude product from 3 replicate batches is quenched with 5% aqueous sodium carbonate, phase-separated, and vacuum distilled at 62–65°C at 10 kPa to obtain benzaldehyde of ≥98% purity. The terminal use is a reactivity probe for impurity profiling in pharmaceutical intermediate synthesis. The benzaldehyde stream is not introduced into food-contact manufacturing without additional purification and compliance with food-grade specification. Strong oxidizers must never be premixed with the bromide promoter before the reactor reaches 120°C. The reactor headspace is inert-purged before air introduction to avoid flammable vapor accumulation.
Non-polar samples such as paraffin waxes, mineral oils, and olefins are insoluble in the standard methanol-based Karl Fischer solvent. A mixed solvent consisting of 20 mL toluene p.a. and 20 mL anhydrous methanol dissolves 0.1–0.5 g of wax before volumetric titration with Hydranal Composite 5. The toluene p.a. lot must show water ≤0.03% by Karl Fischer titration because the method blank is subtracted. A blank value above 0.3 mg water per 20 mL solvent reduces the working range below 10 ppm moisture. The titration cell is maintained at 25°C. Endpoint drift is set to 5 μg/min. High toluene content slows the Karl Fischer reaction rate. A waiting time of 30 s before each determination is used to avoid false endpoint detection. In a volumetric Karl Fischer titrator with a double-platinum electrode and magnetic stirrer, moisture recovery for spiked toluene-wax mixtures at 1.0 mg/g water is expected to fall within 98–102% when the sample is fully dissolved. Undissolved wax aggregates cause surface-adsorbed moisture to titrate slowly, producing low recoveries. This is controlled by heating the solvent mixture to 40°C for samples with melting points above 60°C. The terminal output is a moisture specification report for insulating waxes or mineral oil-based dielectric fluids. Toluene p.a. is used only as a co-solvent, not as the primary Karl Fischer solvent, because pure toluene does not provide sufficient ionic conductivity for reproducible endpoint detection. The measurement follows the volumetric general approach of ISO 760, modified for non-polar sample solubility. Toluene p.a. with evaporation residue above 5 ppm is unsuitable because non-volatile deposits accumulate on the electrode and shift response.
Ultraviolet absorbance profiling of polymer additives, optical brighteners, and stabilizers uses toluene p.a. as a reference solvent because its UV cutoff is located at 285–290 nm. In a double-beam spectrophotometer with 10 mm quartz cuvettes, the solvent baseline is recorded against air and stored. Absorbance measurements are restricted to wavelengths at or above 300 nm. Below this window, intrinsic aromatic absorption reduces linearity and photometric repeatability. The use of 1 mm path-length cuvettes shifts the solvent absorbance floor by approximately one order of magnitude but does not extend the usable cutoff below 285 nm for quantitative work. A typical extract from 0.5 g of polyolefin film dissolved in 50 mL toluene p.a. is scanned from 300 nm to 800 nm. Absorbance bands for hindered amine stabilizers and benzotriazoles are recorded between 310 nm and 380 nm. Method validation follows ASTM E169-16 for ultraviolet-visible quantitative analysis. The terminal output is a lot-specific additive burden report for polymer packaging qualification. Toluene p.a. is not used in sealed cuvettes above 40°C without headspace venting because vapor pressure reaches 5.2 kPa at 30°C and cuvette deformation causes path-length error. Plastic cuvettes are not used with toluene p.a. because aromatic solvent attack produces surface haze and baseline scatter. The solvent blank must be re-recorded after every 10 sample measurements to correct for evaporation-induced concentration drift. Trace benzene and sulfur impurities in lower-grade toluene raise the spectral floor in the 300–320 nm region, which is critical for benzotriazole stabilizer quantitation. Lot-specific certificate absorbance values for Acros Organics toluene p.a. are therefore checked before use in regulated packaging migration studies.
Competitive Toluene, p.a. (pro analysi), Acros Organics prices that fit your budget—flexible terms and customized quotes for every order.
For samples, pricing, or more information, please contact us at +8615651039172 or mail to sales9@boxa-chem.com.
We will respond to you as soon as possible.
Tel: +8615651039172
Email: sales9@boxa-chem.com
Flexible payment, competitive price, premium service - Inquire now!
Acros Organics Toluene, p.a. (pro analysi), supplied under catalogue entries 177480010, 177480025, and 177480250 for 1 L, 2.5 L, and 25 L pack sizes, is an aromatic hydrocarbon solvent with CAS 108-88-3, EC 203-625-9, linear formula C6H5CH3, and molar mass 92.14 g/mol. The product is a clear, colourless liquid with a characteristic aromatic odour. Each lot is accompanied by a certificate of analysis that reports actual values for assay, water, evaporation residue, acidity/alkalinity, and benzene. The pro analysi designation indicates a general analytical reagent grade intended for synthesis, extraction, titration, and preparative work; it is not a pharmacopoeial, GMP, or food-grade material.
Density at 20 °C is 0.866–0.867 g/mL, the boiling range is 110–111 °C, the refractive index n20/D is 1.496–1.497, and the closed-cup flash point is approximately 4 °C. Water solubility is approximately 0.052 g/100 mL at 25 °C. These constants place the material in the low-boiling aromatic solvent class and determine the evaporation and phase-separation behaviour in downstream unit operations. Table 1 summarises characteristic pro analysi thresholds; lot-specific results may be tighter and appear on the certificate of analysis.
| Parameter | Typical p.a. threshold | Test basis |
|---|---|---|
| Assay (GC-FID) | ≥99.5 % | Gas chromatography, area normalisation |
| Water | ≤0.03 % | Karl Fischer titration |
| Evaporation residue | ≤0.001 % | Gravimetric after evaporation at 105–110 °C |
| Acidity/alkalinity | ≤0.0005 meq/g | Acid-base titration after aqueous extraction |
| Benzene | ≤0.05 % | GC-FID or GC-MS external calibration |
| Sulfur compounds (as S) | ≤0.003 % | Oxidative microcoulometry |
| Appearance | Clear, colourless | Visual inspection |
The specification package above differs from industrial toluene primarily in the enforcement of evaporation residue and benzene. For general reagent solvents, test methods are commonly aligned with ISO 6353-2 or supplier-monograph procedures, but the product cannot be considered a harmonised pharmacopoeial monograph. Laboratories using the material for trace metal analysis should verify each lot because p.a. grade does not certify trace metal profiles below the residue specification.
Technical-grade toluene from catalytic reforming or steam cracking may contain paraffinic, olefinic, and sulfur-bearing co-fractions. These species affect downstream reaction selectivity and can leave non-volatile deposits on rotary evaporator vacuum seals, transfer-line fittings, and spray-dryer nozzles. The p.a. grade controls the evaporation residue threshold to ≤0.001 % and reduces benzene to a reported limit of ≤0.05 %. Anhydrous-grade toluene is dried to water values frequently below 0.005 % or 0.001 %, making it preferable for highly moisture-sensitive organometallic complexes. The p.a. grade, with water ≤0.03 %, remains acceptable for routine synthesis but requires molecular-sieve drying when reaction intermediates are intolerant of protic species.
HPLC-grade toluene is controlled for ultraviolet absorbance, fluorescence background, and particulate content because it is used as a mobile-phase component with fixed-wavelength or diode-array detectors. The p.a. grade does not carry those photometric certifications and should not be substituted for HPLC-grade in low-wavelength ultraviolet detection without prior baseline verification. Nitration-grade toluene under ASTM D841-18 is optimised for acid wash colour, paraffin levels, and sulfur constraints relevant to conversion to mono- and di-nitrotoluene. It is not interchangeable with p.a. grade for trace analysis because its evaporation residue and water specifications are less stringent.
Batch-to-batch variance in p.a. toluene is low for the principal parameters, but not zero. In quality-control laboratories, the certificate of analysis is used to adjust Karl Fischer titrant standardisation and gas chromatograph calibration. Where the material is drawn from bulk containers, water uptake can occur through repeated opening under humid ambient conditions. Pre-drying with activated 3A or 4A molecular sieves is expected when relative humidity exceeds 60 %. The product is not supplied as an oxygen-free solvent; dissolved oxygen for oxygen-sensitive catalyst systems can be reduced by sparging with nitrogen or argon.
Organic synthesis campaigns use p.a. toluene as a reaction medium for Grignard reagent formation, lithiation, Friedel-Crafts acylation, and organometallic coupling. In these applications, solvent-borne water can quench an organolithium reagent or reduce catalyst turnover. The limited water threshold of 0.03 % is adequate for many bench-scale procedures, but reactions with preformed organolithium reagents or Schlenk-line manipulations often demand drying to single-digit-ppm levels. In such cases, the p.a. grade is passed through a column of activated alumina or stored over molecular sieves before distillation. Solvent transfer must occur through dried glassware and under inert gas; otherwise, environmental moisture reintroduces proton sources.
For extraction and work-up, the low evaporation residue prevents the accumulation of involatile films during rotary evaporation. A laboratory rotary evaporator operated at 40–60 °C bath temperature and 100–300 mbar vacuum removes toluene efficiently; however, the solvent should not be evaporated to a hard film in the presence of heat-sensitive analytes. Where extractables from polytetrafluoroethylene seals and silicone tubing are a concern, glass or fluoropolymer transfer systems are preferred because the p.a. guarantee does not cover contaminants introduced by external surfaces.
The water-toluene azeotrope, with a boiling point around 84–85 °C, is used in Dean-Stark separations to remove water from esterification and condensation reactions. A Dean-Stark trap or overhead decanter returns the toluene-rich phase to the reactor. This is effective for processes where water is generated in stoichiometric quantities; it is less suitable for gross aqueous mixtures, which are better separated by liquid-liquid extraction before distillation. Emulsions in extraction vessels can be broken by adding anhydrous sodium sulfate or by adjusting ionic strength, but the resulting aqueous phase must be treated as toluene-contaminated waste.
Toluene p.a. is evaluated as a replacement for chlorinated solvents in some cold immersion stripping and alkyd coating removal operations. The substitution is not direct because the vapour is flammable and the Hansen solubility parameters differ from dichloromethane in polar and hydrogen-bonding components. Immersion time must be tested on the actual substrate; prolonged exposure beyond the softening threshold can cause irreversible plasticiser migration in ABS or polycarbonate blends. Published data for this specific configuration is limited. Unlike methylene chloride, p.a. toluene requires explosion-proof mixing equipment and continuous local exhaust ventilation because of the closed-cup flash point of approximately 4 °C.
Polymer and coating laboratories use toluene p.a. to prepare gel permeation chromatography sample solvents for non-polar polymers, dissolve rubber for rheology, and reduce alkyd or polyurethane prepolymer viscosity. High-shear mixers with Cowles blades at tip speeds of 5–15 m/s disperse pigments and fillers into alkyd vehicles; the low residue reduces interference in film defect analysis. However, p.a. grade is not filtered to HPLC standards, so gel permeation chromatography mobile phases may require additional filtration through a 0.45 µm PTFE membrane.
In pharmaceutical intermediate processing, toluene is an ICH Q3C Class 2 solvent with a permitted daily exposure of 8.9 mg/day and a concentration limit of 890 ppm. Benzene is a Class 1 solvent with a far lower permitted daily exposure. Because p.a. toluene reports benzene as a controlled impurity, batch-specific benzene values become relevant when toluene is used near final isolation steps. Residual solvent testing must be conducted on the isolated product because the p.a. certificate does not substitute for batch-specific pharmaceutical release data.
Toluene is classified under the CLP regulation as Flam. Liq. 2 H225, Repr. 2 H361d, Asp. Tox. 1 H304, STOT RE 2 H373, Skin Irrit. 2 H315, and Aquatic Chronic 3 H412. It must be stored away from strong oxidisers, concentrated nitric acid, and strong Lewis acids. Nitration or sulfonation can proceed violently when toluene is mixed with concentrated acid without temperature control; such operations require dedicated reactor design and emergency pressure relief. Static electricity generated during pumping or pouring can ignite the vapour-air mixture; containers and receiving vessels should be electrically bonded and grounded. The lower and upper explosion limits in air are approximately 1.1 % and 7.1 % by volume.
Occupational exposure assessments should apply the relevant national limits. The European Union indicative occupational exposure limit value under Commission Directive (EU) 2017/164 is 192 mg/m³ (50 ppm) as an 8-hour time-weighted average and 384 mg/m³ (100 ppm) as a short-term limit. Engineering controls include local exhaust ventilation, closed-transfer pumps, and vapour recovery units on storage tanks. Where airborne concentrations approach the limit, organic-vapour respirators with AX filters are used only as a secondary control. Spent toluene must be segregated from halogenated solvent waste because mixed recovery streams complicate fractional distillation and may generate corrosive acids during thermal treatment.